FDA Adverse Event
Malfunction
Summary report: N
COAGUCHEK ® XS SYSTEM
MDR report key: 2220358
·
Received August 23, 2011
Report
- Report Number
- 1823260-2011-04489
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 23, 2011
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- GJS
- PMA / PMN Number
- K062925
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. THIS MEDWATCH REPORT IS FOR THE SUSPECT DEVICE USED IN THE COAGUCHEK XS SYSTEM (LOT NUMBER AND EXPIRATION DATE UNKNOWN). REFERENCE MEDWATCH REPORT WITH (B)(6) FOR THE SUSPECT DEVICE USED IN THE COAGUCHEK S SYSTEM.
Description of Event or Problem · 1
CALLER REPORTS THAT DURING A CORRELATION STUDY THE FOLLOWING COAGUCHEK XS/S RESULTS WERE OBTAINED: 2.4 INR/1.9 INR, 3.9 INR/5.5 INR. NO ACTIONS TAKEN BASED ON DEVICE RESULT. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT DEVICE AND REPLACEMENT WAS SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COAGUCHEK ® XS SYSTEM | PROTHROMBIN TIME TEST STRIPS | GJS | ROCHE DIAGNOSTICS | NA | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |