ZIMMER M/L MODULAR FEMORAL STEM
Report
- Report Number
- 1822565-2011-01880
- Event Type
- Injury
- Date Received
- August 16, 2011
- Date of Event
- June 20, 2011
- Report Date
- July 18, 2011
- Manufacturer
- ZIMMER INC
- Product Code
- LPH
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL SUMMARY: THE STERILIZATION PROCESS FOR THESE DEVICES WAS VALIDATED IN ACCORDANCE WITH FDA REGULATIONS AND ISO STANDARDS TO A STERILITY ASSURANCE LEVEL (SAL) OF 1.0 X 10 TO THE POWER OF (-6) OR BETTER. THE MFG LOTS SPECIFIED IN THIS COMPLAINT WERE PROCESSED ACCORDING TO THE VALIDATED STERILIZATION PROCESS PARAMETERS AND MET ALL THE ACCEPTANCE CRITERIA FOR STERILITY RELEASE. THEREFORE, IT IS HIGHLY UNLIKELY THAT THE SPECIFIED DEVICES CAUSED OR CONTRIBUTED TO ANY PT INFECTION. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVAL: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADD'L SUBSTANTIVE INFO BE REC'D, THE COMPLAINT WILL BE REOPENED. ZIMMER, (B)(4) CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THE PT WAS REVISED FOR INFECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZIMMER M/L MODULAR FEMORAL STEM | LPH | ZIMMER INC | 07885059 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention | KINECTIV TECHNOLOGY MODULAR NECK,| CATALOG #00784801200, LOT #60694650 |