FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2220254 · Received August 23, 2011

Report

Report Number
6000001-2011-20301
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
August 15, 2011
Report Date
August 15, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP THAT EXPERIENCED FAILURE CODE 199:117:284:0000 WAS CONFIRMED BY BAXTER PERSONNEL. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS INVOLVED A COLLEAGUE VERSION 1.7 INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 8.11.00. A DEVICE HISTORY REVIEW WAS PERFORMED WITH NO EXCEPTIONS DURING MANUFACTURING FOUND. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THIS PUMP HAS NEVER PREVIOUSLY BEEN SENT TO BAXTER FOR SERVICE AND, THEREFORE, THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE WAS RETURNED TO BAXTER (B)(4) AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(6) SERVICE TECHNICIANS REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 199.117.284.0000. THIS EVENT OCCURRED UPON POWER UP; HOWEVER, IT IS UNKNOWN IN WHICH CARE AREA THIS EVENT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1