RESTORE ULTRA
Report
- Report Number
- 3004209178-2011-06401
- Event Type
- Injury
- Date Received
- August 16, 2011
- Date of Event
- July 13, 2011
- Report Date
- July 21, 2011
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.
IT WAS REPORTED THAT CHARGING OCCURRED MORE THAN EXPECTED. IMPEDANCE MEASUREMENTS WERE WITHIN THE NORMAL RANGE. ADDITIONALLY, THERE WAS A TEMPERATURE WARNING CAME ON AFTER 30 MINUTES OF CHARGING. SETTINGS WERE VERY LOW AND THE RECHARGER ANTENNA WAS REPLACED TWICE. THE PT EXPERIENCED BURNING PAIN AT THE POCKET SITE EVENT WHEN NOT CHARGING. HCP ATTEMPTED TO TREAT THIS BUT FINALLY DECIDED THAT THE PROBLEM COULD ONLY BE RESOLVED WITH REPLACING THE BATTERY. THE BATTERY WAS EXPLANTED (B)(6) 2011 AND WAS REPLACED WITH THE SAME DEVICE MODEL. POSTOPERATIVELY, THE PT WAS GETTING GOOD STIMULATION. THERE HAD NOT BEEN ANY FURTHER CHARGING ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE ULTRA | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37712 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Required Intervention | LEAD: MODEL 3888, LOT# L42354| LOT# NJH703537H| EXTENSION: MODEL 37083, LOT# NKC002708N| PROGRAMMER: MODEL 7434, LOT# YN0011251P| IMPLANTED:| EXPLANTED:| IMPLANTED:| LEAD: MODEL 3890, LOT# J0309488V| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| UNK CONVERSION TYPE: MODEL LEADN, LOT# UNK| EXPLANTED:| EXPLANTED:| EXPLANTED:| IMPLANTABLE NEURO STIMULATOR: MODEL 37711,| EXPLANTED:| IMPLANTED:| EXPLANTED:| PROGRAMMER: MODEL 37742, LOT# NJD045991N| ACCESSORY: MODEL 37752, LOT# NKA029300N| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 37083, LOT# NKC010761N| IMPLANTED:| EXPLANTED:| PROGRAMMER: MODEL 37742, LOT# NJD013259N| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 37083, LOT# NKC002568N| EXPLANTED: |