RESTORE ULTRA
MDR report key: 2220226
·
Received August 16, 2011
Report
- Report Number
- 3004209178-2011-06424
- Event Type
- Injury
- Date Received
- August 16, 2011
- Date of Event
- July 22, 2011
- Report Date
- July 22, 2011
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO., JUNCOS
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): EVAL OF EXTENSION (B)(4) DISCOVERED THAT ALL CONDUCTOR WIRES WERE BROKEN 27.5 CM FROM THE PROXIMAL END AND ALL CIRCUITS WERE OPEN. ANALYSIS OF EXTENSION (B)(4) DISCOVERED THAT ALL CONDUCTOR WIRES WERE BROKEN 19.8 CM FROM THE PROXIMAL END AND ALL CIRCUITS WERE OPEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PT EXPERIENCED A SUDDEN LOSS OF STIMULATION AFTER LIFTING A TIRE. IMPEDANCES ON ALL ELECTRODES WERE MEASURED AT OVER 40,000 OHMS. THE EXTENSIONS WERE REPLACED ON (B)(6) 2011 AND RETURNED TO MFR FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE ULTRA | LGW | MDT PUERTO RICO OPERATIONS CO., JUNCOS | 37712 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR | Required Intervention | EXPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA147852N| IMPLANTED:| LEAD: MODEL 3778, LOT# V498803030| IMPLANTED:| EXPLANTED:| LEAD: MODEL 3778, LOT# V498803032| EXTENSION: MODEL 37081, LOT# NJB087568V| IMPLANTED:| EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE160587N| EXPLANTED:| EXTENSION: MODEL 37081, LOT# NJB082949V |