FDA Adverse Event Injury Summary report: N

RESTORE ULTRA

MDR report key: 2220226 · Received August 16, 2011

Report

Report Number
3004209178-2011-06424
Event Type
Injury
Date Received
August 16, 2011
Date of Event
July 22, 2011
Report Date
July 22, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVAL OF EXTENSION (B)(4) DISCOVERED THAT ALL CONDUCTOR WIRES WERE BROKEN 27.5 CM FROM THE PROXIMAL END AND ALL CIRCUITS WERE OPEN. ANALYSIS OF EXTENSION (B)(4) DISCOVERED THAT ALL CONDUCTOR WIRES WERE BROKEN 19.8 CM FROM THE PROXIMAL END AND ALL CIRCUITS WERE OPEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT EXPERIENCED A SUDDEN LOSS OF STIMULATION AFTER LIFTING A TIRE. IMPEDANCES ON ALL ELECTRODES WERE MEASURED AT OVER 40,000 OHMS. THE EXTENSIONS WERE REPLACED ON (B)(6) 2011 AND RETURNED TO MFR FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ULTRA LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37712 NA

Patients

Seq Age Sex Outcome Treatment
1 23 YR Required Intervention EXPLANTED:| ACCESSORY: MODEL 37752, LOT# NKA147852N| IMPLANTED:| LEAD: MODEL 3778, LOT# V498803030| IMPLANTED:| EXPLANTED:| LEAD: MODEL 3778, LOT# V498803032| EXTENSION: MODEL 37081, LOT# NJB087568V| IMPLANTED:| EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE160587N| EXPLANTED:| EXTENSION: MODEL 37081, LOT# NJB082949V