OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2011-00417
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- July 12, 2011
- Report Date
- July 14, 2011
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE RETURNED PRODUCT WAS EVALUATED AND NO EVIDENCE OF ANY MFG DEFECT WAS DEFECT WAS DETECTED. AN AIR BUBBLE, EQUIVALENT TO 5 UNITS OF INSULIN IN SIZE, WAS FOUND IN THE DEVICE'S INSULIN RESERVOIR. THIS TYPICALLY OCCURS DUE TO THE CUSTOMER INJECTING AIR DURING THE FILL PROCESS, RATHER THAN A MFG PROCESS ISSUE OR PRODUCT DEFECT. USER ERROR IS CONFIRMED AS HAVING CAUSED THE POD TO FAIL TO PERFORM ITS ESSENTIAL FUNCTION. THE OMNIPOD'S USER GUIDE INSTRUCTS USERS ON PROPER FILLING TECHNIQUE AND WARNS TO, "NEVER INJECT AIR INTO THE FILL PORT. DOING SO MAY RESULT IN UNINTENDED OR INTERRUPTED INSULIN DELIVERY." INTERRUPTED INSULIN DELIVERY HAS THE POTENTIAL TO RESULT IN HIGH BLOOD GLUCOSE LEVELS. INSULET HAS INITIATED AN INTERNAL INVESTIGATION TO DETERMINE IF EDUCATION AND TRAINING RELATED TO THIS TOPIC CAN BE IMPROVED. THE INVESTIGATION IS IN-PROCESS AS OF THE DATE OF THIS REPORT.
THE CUSTOMER REPORTED THAT HE WENT TO BED AT 11:14PM WITH A BLOOD GLUCOSE (BG) OF 120MG/DL AND WOKE UP AT 6:20AM WITH A BG OF 185MG/DL. ALTHOUGH THIS IS STILL WITHIN THE "NORMAL" BG RANGE (LESS THAN 250MG/DL) HE STATED THAT HE "HAS NOT WOKEN UP TO A HIGH LIKE THAT BEFORE." HE DID REPORT A BG OF 254MG/DL THE DAY BEFORE, BUT HE IS "LESS CONCERNED ABOUT THE HIGHS AND LOWS DURING THE DAY AND MORE CONCERNED ABOUT THE CLIMB IN HIS BG'S OVERNIGHT FROM 11:14PM TO 6:20PM." HE "BUMPED" THE POD BED AND WANTS TO MAKE SURE THE DEVICE IS FUNCTIONING PROPERLY. THE PRODUCT WILL BE RETURNED FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |