FDA Adverse Event Malfunction Summary report: N

THERMOPHORE ARTHRITIS PAD

MDR report key: 2220177 · Received August 15, 2011

Report

Report Number
1811605-2011-00097
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
June 1, 2011
Report Date
June 16, 2011
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE SWITCH FAILED. UPON EXAMINATION, WE DISCOVERED THAT A PCB COMPONENT APPEARS TO HAVE SHORTED, CAUSING OTHER COMPONENT TO FAIL. THE FLAME-RETARDANT SWITCH HOUSING CONTAINED THE EVENT PROPERLY, BUT WAS DEFORMED IN THE PROCESS. AT OUR REQUEST, THE MFR OF THE SWITCH HAS ENACTED SEVERAL CAPA PROJECTS TO IMPROVED THE QUALITY OF THE SWITCH. THE OCCURRENCE OF BOARD COMPONENT FAILURES HAS DROPPED SIGNIFICANTLY IN THE PAST 12 MONTHS.

Description of Event or Problem · 1

IT WAS REPORTED THE SWITCH FAILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE ARTHRITIS PAD ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 155

Patients

Seq Age Sex Outcome Treatment
1