FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 2220176
·
Received August 15, 2011
Report
- Report Number
- 1811605-2011-00094
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 23, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THE UNIT STOPPED WORKING. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK PREVENTS THE WIRES IN THE CORD FROM MAKING CONTINUOUS CONTACT AND COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE BENDING. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND THE CORD NEAR THE SWITCH, WE ARE TAKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.
Description of Event or Problem · 1
IT WAS REPORTED THE UNIT STOPPED WORKING.