FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 2220176 · Received August 15, 2011

Report

Report Number
1811605-2011-00094
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
June 1, 2011
Report Date
June 23, 2011
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE UNIT STOPPED WORKING. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK PREVENTS THE WIRES IN THE CORD FROM MAKING CONTINUOUS CONTACT AND COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE BENDING. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND THE CORD NEAR THE SWITCH, WE ARE TAKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.

Description of Event or Problem · 1

IT WAS REPORTED THE UNIT STOPPED WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 055

Patients

Seq Age Sex Outcome Treatment
1