FDA Adverse Event Malfunction Summary report: N

THERMOPHORE ARTHRITIS PAD

MDR report key: 2220175 · Received August 15, 2011

Report

Report Number
1811605-2011-00095
Event Type
Malfunction
Date Received
August 15, 2011
Date of Event
June 1, 2011
Report Date
June 23, 2011
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE USER REPORTED SPARKS FROM THE SWITCH WHEN UNPLUGGING IT FROM THE OUTLET. UPON EXAMINATION, THE CORD APPEARS TO HAVE BEEN CUT BY SOME UNUSUAL EVENT PRIOR TO THE FAILURE. WE ARE UNABLE TO DETERMINE WHETHER THE FAILURE WOULD HAVE OCCURRED HAD THE CORD NOT BEEN DAMAGED. THE CORD APPEARS TO HAVE BEEN CAUGHT IN SOMETHING AND CUT BEFORE THE EVENT OCCURRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORD EMITTED SPARKS NEAR THE SWITCH WHEN THE USER UNPLUGGED THE UNIT FROM THE OUTLET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE ARTHRITIS PAD ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 156

Patients

Seq Age Sex Outcome Treatment
1