THERMOPHORE AUTOMATIC MOIST HEAT PACK
Report
- Report Number
- 1811605-2011-00096
- Event Type
- Malfunction
- Date Received
- August 15, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 21, 2011
- Manufacturer
- BATTLE CREEK EQUIPMENT CO.
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT WAS REPORTED THE UNIT STOPPED WORKING. EXAMINATION OF THE UNIT REVEALED A BREAK IN THE CORD AT THE ENTRANCE TO THE STRAIN RELIEF AREA NEAR THE SWITCH. THE BREAK PREVENTS THE WIRES IN THE CORD FROM MAKING CONTINUOUS CONTACT AND COULD POSSIBLY EXPOSE BARE WIRES WHEN THE CORD IS BENT. THE FAILURE MAY BE A RESULT OF MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS ON THE PART OF THE CONSUMER BY SUBJECTING THE CORD TO EXCESSIVE BENDING. THE CONDITION OF THE PAD SUGGESTS IT WAS USED EXCESSIVELY AND INCORRECTLY. ALTHOUGH USERS ARE INSTRUCTED NOT TO BEND THE CORD NEAR THE SWITCH, WE ARE TAKING APPROPRIATE ACTION TO MAKE THE DESIGN MORE ROBUST BY INITIATING A CAPA PROJECT ((B)(4)) TO PREVENT THIS FAILURE FROM RECURRING.
IT WAS REPORTED THE UNIT STOPPED WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT CO. | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |