FDA Adverse Event Injury Summary report: N

ION

MDR report key: 22201498 · Received June 13, 2025

Report

Report Number
2955842-2025-25251
Event Type
Injury
Date Received
June 13, 2025
Date of Event
May 22, 2025
Report Date
May 22, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
EOQ
UDI-DI
00886874116234
PMA / PMN Number
K182188
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THERE WAS NO ALLEGATION THAT A MALFUNCTION OF AN ION SYSTEM, INSTRUMENT, OR ACCESSORY OCCURRED. A SYSTEM LOG REVIEW COULD NOT BE PERFORMED BECAUSE THE SYSTEM LOGS WERE NOT AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING AN ION ENDOLUMINAL LUNG BIOPSY PROCEDURE THE PATIENT DEVELOPED A PNEUMOTHORAX REQUIRING A CHEST TUBE AND HOSPITALIZATION. A CONE BEAM WITH CIOS SPIN WAS PERFORMED AFTER SUCCESSFULLY NAVIGATED TO THE NODULE. WHEN THE SCAN WAS COMPLETED, THE PHYSICIAN REVIEWED IT AND NOTICED A PNEUMOTHORAX. THE PHYSICIAN PROCEEDED WITH THE BIOPSY SINCE HE WAS AT THE RIGHT SPOT. AFTER FINISHING THE ION PORTION OF THE PROCEDURE, A CHEST TUBE WAS PLACED IN THE PATIENT DURING THE SAME PROCEDURE. THE BIOPSY WAS APPROXIMATELY 20 MINUTES AND THERE WAS A DELAY DUE TO THE CHEST TUBE PLACEMENT. THE PATIENT WAS ADMITTED TO THE HOSPITAL AND WAS LIKELY TO STAY THE NIGHT. THE PHYSICIAN CONFIRMED THAT THE ION SYSTEM DID NOT CAUSE THE PNEUMOTHORAX, SINCE THE PATIENT WAS VERY SICK AND HAD A LOT OF EMPHYSEMA WITH BULLAE. THERE WAS NO ALLEGATION THAT A MALFUNCTION OF AN ION SYSTEM, INSTRUMENT, OR ACCESSORY OCCURRED. INTUITIVE SURGICAL INC., (ISI) HAS MADE MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT. HOWEVER, THERE WAS NO RESPONSE RECEIVED FROM THE PHYSICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
376891 ION SYSTEM CART EOQ INTUITIVE SURGICAL, INC 380748-65 N/A 00886874116234

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R ION ENDOLUMINAL SYSTEM.