YPSOPUMP INSET II
Report
- Report Number
- 3003442380-2025-10589
- Event Type
- Malfunction
- Date Received
- June 13, 2025
- Date of Event
- June 2, 2025
- Report Date
- June 17, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED. THE REFERENCE SAMPLES FOR THE LOT 6010360 HAVE BEEN TESTED IN DATABASE 2133074 ON 13-MAR-2025. TEST RESULTS: VISUAL TEST ACCORDING TO WORK INSTRUCTION (WI) VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR FLOW ACCORDING TO WI VERSION 1 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010360 WAS MANUFACTURED ACCORDING TO THE WI VERSION 75 MANUFACTURED IN THE LINE 6, ON 23/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED; NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 16-JUN-2025 AGAINST MALFUNCTION CODE LEAKAGE (CUSTOMER STATES INFUSION SET LEAKS) (SOURCE/CAUSE CANNOT BE IDENTIFIED, ONLY USE WHEN A SPECIFIC MALFUNCTION CANNOT BE DETERMINED) AND LOT 6010360, WITH NO TREND IDENTIFIED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
H11: SINCE NO LOT NUMBER IS AVAILABLE, A DETAILED INVESTIGATION, TESTS OR REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONCLUDED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH PMS PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WILL FLAG ON THE TRIPS AND ALERTS ACCORDING TO THE OMQR PROCEDURE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN SWITZERLAND. IT WAS REPORTED THAT PATIENT FACED INFUSION SET LEAKING ON (B)(6) 2025. THE INSERTION SITE WAS THIGH. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 552253 | YPSOPUMP INSET II | UNO INSET II 60/6 GREY Y-CAP 10PAK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 86-080-52B9 | 6010360 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |