TRIAGE TOX DRUG SCREEN WITH MTD
Report
- Report Number
- 2027969-2011-01531
- Event Type
- Malfunction
- Date Received
- July 13, 2011
- Date of Event
- June 14, 2011
- Report Date
- July 13, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- DJR
- PMA / PMN Number
- 060791
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
PRODUCT SUPPORT TESTED 3 REPLICATES IN QA RETAINED W49252 DEVICES WITH DRUG FREE URINE, TOX 1.25X CALIBRATOR AND R&D NEGATIVE URINE SAMPLE. RESULTS AS FOLLOWS: NO PHYSICAL DEFECTS OR ERROR CODES OBSERVED ON DEVICES. TIME TO RESULTS: WITHIN 15 MINUTES. GLOBAL STATPAK TRACES: ELEVATED BASELINE OBSERVED ON EACH IN-HOUSE CONTROL TESTED. NO DISCREPANT RESULTS OF FALSE NEGATIVE APAP OBSERVED. DEVICES PERFORMED WITHIN MFG SPECIFICATION. PRODUCT SUPPORT ALSO TESTED 2 REPLICATES OF RETURNED URINE ON QC RETAINED W49252 DEVICES. THE RETURNED URINE WAS ALSO TESTED ON ADULTERATION STRIPS AND SENT OUT TO NMS LABS FOR FURTHER HPLC ANALYSIS. RESULTS AS FOLLOWS: ONE RETURNED URINE LABELED: (B)(6). NO SAMPLE LEFT AFTER TESTING COMPLETED. PH 10, SPECIFIC GRAVITY 1.007 UG. TOX W49252, 2 REPS TESTED: ALL POSITIVE FOR APAP, OPI, BZO. NO OTHER POSITIVE RESULTS OBSERVED. TIME TO RESULTS WITHIN 15 MINUTES. TRIAGE TOX THRESHOLD FOR APAP: 5 UG/ML. NO ERROR CODES REPORTED AND NO PHYSICAL DEFECTS OBSERVED ON DEVICES. CUSTOMER'S NEGATIVE APAP RESULTS WERE NOT REPRODUCED. RETURNED URINE SAMPLE TESTED POSITIVE ON TOX W49252 AND HPLC DETECTED 75 UG/ML APAP WHICH IS ABOVE TOX ASSAY THRESHOLD OF 5 UG/ML. POSITIVE OPI WAS DUE TO VICODIN MEDICATION. POSITIVE BZO MIGHT ALSO BE DUE TO MEDICATION. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CUSTOMER REPORTED FALSE NEGATIVE ACETAMINOPHEN (APAP) FOR ONE PT TESTED ON THE URINE TOX SCREEN TEST. PT TAKES 750 MG OF TYLENOL, CHRONICALLY, THREE TIMES PER DAY. SHE WAS ALSO TAKING VICODIN. THREE TESTS SINCE APRIL, THE OPI WAS ALWAYS POSITIVE, BUT APAP WAS ALWAYS NEGATIVE. PHYSICIAN IS QUESTIONING THE RESULT OF TOX. OTHER PT TESTS GAVE NORMAL POSITIVE APAP RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRIAGE TOX DRUG SCREEN WITH MTD | URINE DRUG SCREEN TEST | DJR | ALERE SAN DIEGO, INC. | 94400 | W49252 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |