ARCTIC SUN® 5000
Report
- Report Number
- 1018233-2025-04517
- Event Type
- Malfunction
- Date Received
- June 12, 2025
- Date of Event
- April 7, 2025
- Report Date
- June 11, 2025
- Manufacturer
- MEDIVANCE, INC. ¿ 1725056
- Product Code
- DWJ
- UDI-DI
- 00801741080142
- PMA / PMN Number
- K161602
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE REPORTED ISSUE WAS CONFIRMED. THE ROOT CAUSE OF THE FAILED CALIBRATION WAS DETERMINED TO BE A FAILED CIRCULATION PUMP. THE DEVICE WAS EVALUATED UPON RECEIPT. DURING A CALIBRATION CHECK T3 WOULD NOT READ BELOW 30 DEGREES C DUE TO NO FLOW THROUGH THE MANIFOLD. CALIBRATION ERROR 80 RECEIVED. SERVICED THE CIRCULATION PUMP. MIXING PUMP MOUNT WAS FOUND TO BE FRACTURED. FLOWMETER WAS READING OVER 300 ML/M LOWER THAN THE CTU VALUE. A 2000-HOUR PM WAS COMPLETED ON THE DEVICE. REPLACED MIXING PUMP MOTOR. REPLACED FLOWMETER. AS PART OF THE PM, SERVICED MIXING PUMP, REPLACED HEATER, DRAIN VALVES, BOTH MANIFOLD O-RINGS, DOUBLE BEND TUBE, AND CHILLER EVAPORATOR OUTLET TUBE. REPLACED THE CONTROL PANEL COIN CELL DUE TO AGE. THE ARCTIC SUN 5000 PASSED ALL PERFORMANCE TESTING, CALIBRATION AND ELECTRICAL SAFETY TESTS AND IS FUNCTIONING PROPERLY AND READY FOR USE. THE COMPLAINT OR REPORTED ISSUE WAS CONFIRMED THROUGH OTHER ELEMENTS OF THE INVESTIGATION TO NOT BE LABELING OR PACKAGING RELATED. A DHR REVIEW IS NOT REQUIRED AS THE REPORTED EVENT IS NOT AN OUT OF BOX FAILURE AND THEREFORE THE REPORTED EVENT IS NOT MANUFACTURING RELATED. UDI FORMAT UPDATED WITH AVAILABLE PRODUCT INFORMATION PER GUDID. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
IT WAS REPORTED THAT THE BIOMED STATED THAT THE ARCTIC SUN DEVICE WAS BEING CALIBRATED AND RECEIVED AN ERROR 82 (WATER CHECK TIME OUT - OTHER CONDITION). PER ADDITIONAL INFORMATION UPDATED FROM TTM ON 07APR2025, BIOMED STATED ON LAST STEP OF CALIBRATION THEY KEEP GETTING ERROR 82 (WATER CHECK TIME OUT - OTHER CONDITION). PER SAMPLE EVALUATION RESULTS RECEIVED VIA TASK ON 06JUN2025, IT WAS REPORTED THAT THE MIXING PUMP MOUNT WAS FOUND TO BE FRACTURED. FLOWMETER WAS READING OVER 300 ML/M LOWER THAN THE CTU VALUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2200707 | ARCTIC SUN® 5000 | ARCTIC SUN DEVICE | DWJ | MEDIVANCE, INC. ¿ 1725056 | NA | 00801741080142 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Other |