FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 22189363 · Received June 11, 2025

Report

Report Number
2955842-2025-24466
Event Type
Malfunction
Date Received
June 11, 2025
Date of Event
May 21, 2025
Report Date
May 21, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED FIELDS: H2, H3, ANNEX CODES C, D. INTUITIVE SURGICAL, INC. RECEIVED THE MASTER TOOL MANIPULATOR (MTM) TO PERFORM FAILURE ANALYSIS HOWEVER, THE CUSTOMER'S REPORTED EVENT COULD NOT BE REPLICATED. UPON VISUAL INSPECTION, NO ISSUES WERE FOUND THAT WOULD BE RELATED TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. BASED ON THE FIELD EVALUATION, THIS REPORTED EVENT WAS CONFIRMED. THE FSE REPLACED THE MTM ON THE SURGEON SIDE CONSOLE 2 (SSC2) TO RESOLVE THE CUSTOMERS REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE MASTER TOOL MANIPULATOR (MTM) PRODUCT HAS BEEN RECEIVED; HOWEVER, FAILURE ANALYSIS HAS NOT BEEN COMPLETED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER REPORTED THAT UNIVERSAL SURGICAL MANIPULATOR (USM) 3 STOPPED WORKING. THE CUSTOMER RESTARTED THE SYSTEM AND WAS OBSERVING ERRORS POINTING TO THE RIGHT MASTER TOOL MANIPULATOR (MTM) ON THE SECOND SURGEON SIDE CONSOLE (SSC). TECHNICAL SERVICE ENGINEERING (TSE) CONFIRMED THE ERROR. THE PROCEDURE WAS CONTINUED ON A SECONDARY SSC WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103540 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-29 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES