METRX® SYSTEM
Report
- Report Number
- 1030489-2025-02205
- Event Type
- Malfunction
- Date Received
- June 11, 2025
- Date of Event
- May 22, 2025
- Report Date
- October 13, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HRX
- UDI-DI
- 00885074210353
- PMA / PMN Number
- K002931
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
E1: FIRST NAME AND LAST NAME OF INITIAL REPORTER IS UNKNOWN. G2: COUNTRY OF ORIGIN IS JAPAN. H3 - EVALUATION OF THE RETURNED DEVICE WAS NOT COMPLETED AT THE TIME OF THIS REPORT. A FOLLOW UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
D4:UDI UPDATED. H3:PRODUCT ANALYSIS FOR PRODUCT:9560100, LOTNO:1192778. DURING THE INSPECTION AT THE TIME OF ACCEPTANCE, THE REQUESTED SYMPTOMS WERE DUPLICATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THERE IS IMAGING ABNORMALITY, LENS VISUAL INSPECTION REVEALED THAT THE INSIDE OF THE LENS HAD BECOME DIRTY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 94336 | METRX® SYSTEM | ARTHROSCOPE | HRX | MEDTRONIC SOFAMOR DANEK USA, INC | 9560100 | 1192778 | 00885074210353 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |