Report

Report Number
1030489-2025-02205
Event Type
Malfunction
Date Received
June 11, 2025
Date of Event
May 22, 2025
Report Date
October 13, 2025
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HRX
UDI-DI
00885074210353
PMA / PMN Number
K002931
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: FIRST NAME AND LAST NAME OF INITIAL REPORTER IS UNKNOWN. G2: COUNTRY OF ORIGIN IS JAPAN. H3 - EVALUATION OF THE RETURNED DEVICE WAS NOT COMPLETED AT THE TIME OF THIS REPORT. A FOLLOW UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

D4:UDI UPDATED. H3:PRODUCT ANALYSIS FOR PRODUCT:9560100, LOTNO:1192778. DURING THE INSPECTION AT THE TIME OF ACCEPTANCE, THE REQUESTED SYMPTOMS WERE DUPLICATED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THERE IS IMAGING ABNORMALITY, LENS VISUAL INSPECTION REVEALED THAT THE INSIDE OF THE LENS HAD BECOME DIRTY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94336 METRX® SYSTEM ARTHROSCOPE HRX MEDTRONIC SOFAMOR DANEK USA, INC 9560100 1192778 00885074210353

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown