FDA Adverse Event Injury Summary report: N

MINIMED MIO ADVANCE

MDR report key: 22181725 · Received June 11, 2025

Report

Report Number
8021545-2025-01479
Event Type
Injury
Date Received
June 11, 2025
Date of Event
May 23, 2025
Report Date
October 3, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018877
PMA / PMN Number
K173879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED STATES. SINCE NO LOT NUMBER IS AVAILABLE, A DETAILED INVESTIGATION, TESTS ON REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONDUCTED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH PMS PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WOULD FLAG THE TRIPS AND ALERTS ACCORDING TO THE OMQR PROCEDURE.

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (8021545-2025-01479), WAS SUBMITTED ON 11-JUN-2025. UPON RECEIVING ADDITIONAL INFORMATION, IT WAS DISCOVERED THAT MANUFACTURING DATE HAS UPDATED TO 24-FEB-2025. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H4: MANUFACTURING DATE. H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011775. IN QUESTION WAS MANUFACTURED AT THE OSTED SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 03-OCT-2025AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH" , "LOT NUMBER" CRITERIA EQUAL 6011775. THE COUNT OF COMPLAINT IS 0 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE 6011775 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) [80] APPENDIX 1 BATCHCARD FOR PRODUCTION OF PACKAGING ROOM ON 24-FEB-2025 WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PRODUCT WAS RETURNED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT WENT TO AN EMERGENCY ROOM (ER) AND EVENTUALLY GOT HOSPITALIZED ON (B)(6)2025 DUE TO HYPOGLYCEMIA EVENT. BLOOD GLUCOSE LEVEL WAS 23 MG/DL AT THE TIME OF THE EVENT. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE, ADDITIONAL PATIENT OR EVENT DETAILS WERE RECEIVED WHICH HAVE BEEN ADDED IN H11.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103077 MINIMED MIO ADVANCE UNO MIO ADV. GREY 60/6 SC1 10-PK INT FPA UNOMEDICAL A/S MMT-242A UNKNOWN 05705244018877

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown