FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2218132 · Received August 22, 2011

Report

Report Number
2050012-2011-04658
Event Type
Malfunction
Date Received
August 22, 2011
Date of Event
July 22, 2011
Report Date
July 22, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4)-2011 A BEC FIELD SERVICE ENGINEER (FSE) NOTED THE EIC AND ION-SELECTIVE ELECTRODE (ISE) DRAIN WERE OVERFLOWING INDICATING AN ISSUE WITH THE ISE VACUUM. FSE ENABLED THE MODULAR CHEMISTRY SIDE AND HOMED. THE ISE DRAIN HAD BEEN FULL AND IT DRAINED. FSE PRIMED 20X WITH NO OVERFLOWS. FSE CHECKED EIC, VACUUM DRAIN LINES AND ISE DAMPER FOR BLOCKAGES AND FOUND NONE. PRIMED 250 ML OF 25% WARM BLEACH THRU THE VACUUM DRAIN LINE. CALIBRATED AND RAN PRECISIONS TO TEST. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT THAT THE ELECTROLYTE INJECTION CUP (EIC) OVERFLOWED ON THE UNICEL DXC 800P SYNCHRON CHEMISTRY ANALYZER. PER CUSTOMER, NO PATIENT RESULTS HAVE BEEN AFFECTED BY THIS EVENT. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1