STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-09597
- Event Type
- Injury
- Date Received
- June 10, 2025
- Date of Event
- April 28, 2025
- Report Date
- August 6, 2025
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON JUNE 24, 2025, WITH LOT NUMBER 2082750. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: DEFLATION: OBSERVED AN OPENING ASSESSED AS FOLD FLAW OPENING AS PER THE INVESTIGATION PROCEDURE, CREASES, WEAR ABRASION WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 91071 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2082750 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Female | Required Intervention |