FDA Adverse Event Injury Summary report: N

FDR GO

MDR report key: 22174295 · Received June 10, 2025

Report

Report Number
1000513161-2025-00020
Event Type
Injury
Date Received
June 10, 2025
Date of Event
May 20, 2025
Report Date
June 9, 2025
Manufacturer
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Product Code
IZL
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

REF. COMP # (B)(4).

Description of Event or Problem · 0

ON (B)(6) 2025, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING FDR GO PORTABLE X-RAY SYSTEM. ON (B)(6) 2025, A FUJIFILM ENGINEER WAS SERVICING THE FDR GO FOR A BATTERY CHARGING ISSUE. AN ELECTRICAL ARC WAS DISCHARGED WHILE SWAPPING OUT A CIRCUIT BOARD. THE ENGINEER SUSTAINED 1ST, 2ND, AND 3RD DEGREE BURNS TO THE RIGHT HAND (INDEX AND MIDDLE FINGER) WHICH REQUIRED DEBRIDING AND OVERNIGHT HOSPITALIZATION FOR OBSERVATION. ON (B)(6) 2025, FUJIFILM TEAM ARRIVED AT THE CUSTOMER SITE TO INVESTIGATE THE INCIDENT, REVIEW SERVICING PRACTICES, ASSIST WITH TAKING APART THE FDR GO AND IMPLEMENT CORRECTIVE ACTIONS. THE INJURED ENGINEER STATED HE WAS REPLACING A POWER BOARD WHEN IT EXPLODED IN HIS HANDS. A FUJIFILM COWORKER HEARD A LOUD POP AND SAW THE INJURED ENGINEER ON THE GROUND WITH BURNS TO THEIR HAND. BOTH FUJIFILM EMPLOYEES STATED SERVICE WORK WAS PERFORMED ACCORDING TO FUJIFILM FDR GO SERVICE PROCEDURE AND IT WAS UNPLUGGED DURING SERVICING. ON (B)(6) 2025, ANOTHER FUJIFILM TEAM VISITED THE SITE TO ASSIST WITH CONTINUED INVESTIGATION OF THE POTENTIAL CAUSES. THE TEAM OBSERVED SIGNIFICANT BURN MARKS ON THE CIRCUIT BOARD. PHOTOS WERE TAKEN FROM ALL ANGLES. DISASSEMBLY THE FDR GO AND CONTINUITY CHECKS WHEN SYSTEM WAS TURNED ON OR OFF. THIS INCLUDED BATTERY VOLTAGE CHECKS, BREAKER INSPECTION, AND TESTS FOR ELECTRICAL SHORTS. ALL FUSES AND CAPACITORS WERE VERIFIED TO BE INTACT. THIS REPORT IS BEING SUBMITTED DUE TO THE CONFIRMED INJURY TO THE SERVICE ENGINEER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1318330 FDR GO SYSTEM, X-RAY, MOBILE IZL SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION FDR GO UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other