DUKE UNIV DURHAM NC 1
Report
- Report Number
- 1718850-2011-00082
- Event Type
- Other
- Date Received
- July 15, 2011
- Date of Event
- April 7, 2011
- Report Date
- June 15, 2011
- Manufacturer
- SORIN GROUP USA, INC.
- Product Code
- DWF
- PMA / PMN Number
- PREAMENDMENT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
PT INFORMATION WAS NOT PROVIDED. SORIN GROUP (B)(4) MANUFACTURES THE CUSTOM HEART LUNG PACK. THE REVOLUTION IS A COMPONENT OF THE PACK AND IS MANUFACTURED BY SORIN GROUP (B)(4). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4)RECEIVED A REPORT OF A SQUEALING NOISE FROM THE REVOLUTION PUMP HEAD AND PUMP DRIVER, WHICH RESULTED IN LOW FLOW. THE PUMP DRIVER WAS MANUALLY CRANKED TO WEAN THE PT FROM BYPASS, THEN THE PUMP HEAD WAS RESEATED INTO THE PUMP DRIVER AND THE CASE CONTINUED. THERE WAS NO REPORT OF PT INJURY. HOWEVER, FURTHER COMMUNICATION WITH THE PERFUSIONIST REVEALED THAT SUB-OPTIMAL HAND CRANKING WAS NOT ABLE TO GENERATE SUFFICIENT FLOW FOR APPROXIMATELY 4 MINUTES BEFORE ASSISTANCE WAS PROVIDED. SORIN GROUP (B)(4) MANUFACTURED THE REVOLUTION PUMP HEAD INVOLVED IN THIS INCIDENT. THE PUMP HEAD IS A COMPONENT OF THE CUSTOMER PERFUSION PACK. THE 510(K) NUMBER OF THE REVOLUTION PUMP HEAD IS K030462. THE REVOLUTION PUMP HEAD WAS RETURNED TO SORIN GROUP (B)(4) FOR EVALUATION. NO DEFECTS OR ABNORMALITIES WERE NOTED DURING VISUAL INSPECTION. THE RETURNED UNIT WAS SET UP IN A CIRCUIT AND RUN FOR SIX HOURS WHILE LISTENING FOR ABNORMAL NOISE AND MONITORING THE FLOW RATE. THE UNIT PERFORMED AS WITHOUT ANY ISSUE. THE FLOW RATE WAS MAINTAINED AND THE PUMP HEAD DID NOT PRODUCE ABNORMAL NOISE. THE REPORT OF HEARING A SQUEALING NOISE AND NOTICING THE FLOWS DROPPING OFF COULD NOT BE REPRODUCED AND THE CAUSE FOR THE CLINICIANS CONCERN COULD NOT BE DETERMINED. NO STRUCTURAL OR PERFORMANCE DEFECTS WERE FOUND. NO FURTHER ACTION IS DEEMED NECESSARY.
SORIN GROUP (B)(4) RECEIVED A REPORT OF A SQUEALING NOISE FROM THE REVOLUTION PUMP HEAD AND PUMP DRIVER, WHICH RESULTED IN LOW FLOW. THE PUMP DRIVER WAS MANUALLY CRANKED TO WEAN THE PT FROM BYPASS, THEN THE PUMP HEAD WAS RESEATED INTO THE PUMP DRIVER AND THE CASE CONTINUED. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DUKE UNIV DURHAM NC 1 | CUSTOM PERFUSION PACK | DWF | SORIN GROUP USA, INC. | NA | 1103100060 |