ENDOWRIST
Report
- Report Number
- 2955842-2025-24362
- Event Type
- Malfunction
- Date Received
- June 9, 2025
- Date of Event
- May 16, 2025
- Report Date
- May 20, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874119808
- PMA / PMN Number
- K214095
- Removal / Correction Number
- ISIFA2024-09-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CI
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
AN INVESTIGATION IS IN PROGRESS TO DETERMINE THE CAUSE AN RMA WAS ISSUED TO EVALUATE THE INTUITIVE SURGICAL, INC. (ISI) DEVICE. OF THIS REPORTED EVENT. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE.
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A BROKEN GRIP CABLE. WHEN CABLE BREAKAGE OCCURS, THE INSTRUMENT IS IMMEDIATELY INOPERABLE DUE TO LOSS OF FUNCTIONALITY. THE INSTRUMENT HAS TUNGSTEN DRIVE CABLES TO TRANSMIT MOTION FROM THE INPUT DISCS IN THE HOUSING, THROUGH THE MAIN SHAFT, AND TO THE DISTAL INSTRUMENT TIP. THESE CABLES ARE EXPOSED AT THE INSTRUMENT TIP AND CONTROL MOVEMENTS AT THE WRIST. THE COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS, WHICH INDICATES THAT THE DEVICE MAY HAVE CONTRIBUTED TO THE CUSTOMER REPORTED ISSUE. THE PROBABLE ROOT CAUSE OF CABLE BREAKAGE, WHETHER PARTIAL OR FULL, IS ATTRIBUTED TO DAMAGE TO THE CABLE THROUGH EXTERNAL COLLISIONS, DURING USE, OR DURING REPROCESSING. THIS ISSUE CAN BE RESOLVED BY USING AN ALTERNATE INSTRUMENT TO COMPLETE THE PROCEDURE.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSTRUMENT STOPPED CLOSING ONE OF ITS JAWS, SO IT WAS REMOVED FOR EXAMINATION. IT WAS OBSERVED THAT ONE OF THE PULLEY CABLES WAS CUT, SO IT WAS REPLACED WITH A BACKUP INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1632484 | ENDOWRIST | FENESTRATED BIPOLAR FORCEPS | NAY | INTUITIVE SURGICAL, INC | 471205-17 | K11241009 0001 | 00886874119808 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |