DA VINCI SP
Report
- Report Number
- 2955842-2025-24314
- Event Type
- Malfunction
- Date Received
- June 9, 2025
- Date of Event
- May 13, 2025
- Report Date
- May 13, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874114605
- PMA / PMN Number
- K182371
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE PROBABLE ROOT CAUSE IS ATTRIBUTED TO MECHANICAL INCONSISTENCY AT THE STERILE ADAPTER (SA) MOUNT INTERFACE ON THE PATIENT SIDE MANIPULATOR (PSM), WHICH PREVENTED PROPER SA RETENTION DURING INSTALLATION. THIS MISALIGNMENT LIKELY RESULTED IN THE DRAPE NOT SECURELY ATTACHING TO THE MANIPULATOR. SINCE NO FUNCTIONAL ANOMALIES WERE OBSERVED DURING INSTRUMENT CONTROL OR DRIVE TESTS, THE SECOND COMPLAINT COULD NOT BE SUBSTANTIATED.
AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE FOUND THE DRAPE WAS NOT HOLDING ONTO THE PATIENT SIDE MANIPULATOR (PSM). THE FSE REPLACED THE PSM TO RESOLVE THE ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE THE PSM TO PERFORM FAILURE ANALYSIS (FA). FA WAS ABLE TO CONFIRM ISSUE OF DRAPE NOT HOLDING ONTO PSM AND REPRODUCE THE ISSUE DURING IN-HOUSE TESTING. FA WAS NOT ABLE TO CONFIRM NOR REPRODUCE THE ISSUE OF SURGEON COULD NOT TAKE CONTROL OF THE INSTRUMENTS. DURING VISUAL INSPECTION, NO ISSUES WERE FOUND THAT WOULD BE RELATED TO THE REPORTED EVENT. THE UNIT WAS INSTALLED ONTO A GOLDEN SYSTEM AND WHEN INSTALLING A STERILE ADAPTER (SA), THE SA WOULD POP OFF ON ONE SIDE. DESPITE THE RETENTION ISSUES, FA WAS ABLE TO INSTALL INSTRUMENTS AND DRIVE THE PSM WITH NO ISSUES. THE UNIT WAS THEN INSTALLED ONTO A PSC FIXTURE TEST PLATFORM (PFTP) WHERE IT PASSED ALL TESTING WITHIN SPECIFICATION. AS A RESULT OF THESE FINDINGS, FAILURE ANALYSIS HAS CONCLUDED THAT THE SA MOUNT IS THE ROOT CAUSE OF THIS ISSUE. NOTE: THE UNIT PASSED THE OLED TEST, SERVO TEST, PRELOAD MOTOR HEALTH CHECK, BRAKE SPEC TEST, BRAKE REPEATABILITY TEST, CLUTCH BUTTON CHECK, SA PLUNGER CHECK, SA ENGAGEMENT CHECK, PRELOAD SENSOR CHECK, INSTRUMENT SENSOR CHECK, INSTRUMENT ENGAGE SPLINE, SINE CYCLE, SMOOTHNESS TEST, FRICTION TEST, FRICTION HIGH-SPEED SWEEP, AND BACKLASH TEST.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SINGLE PORT SURGICAL PROCEDURE, THE CUSTOMER REPORTED TO INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SERVICE ENGINEER (TSE) THAT THE SURGEON SIDE CONSOLE (SSC) COULD NOT TAKE CONTROL. THE CUSTOMER POWER CYCLED THE SYSTEM TO RESOLVE THE ISSUE; HOWEVER, THE SURGEON STILL COULD NOT TAKE CONTROL OF THE INSTRUMENTS. TSE CONFIRMED ON SYSTEM LOGS NORMAL OPERATION AND INDICATED SURGEON AT THE CONSOLE VIEWER AND ARMS WERE CORRECTLY ASSIGNED WITH NO ERRORS. THE PROCEDURE WAS CONVERTED TO ANOTHER SYSTEM WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1654561 | DA VINCI SP | PATIENT SIDE CART | NAY | INTUITIVE SURGICAL, INC | 380601-40 | N/A | 00886874114605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |