FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 22155115 · Received June 5, 2025

Report

Report Number
1723170-2025-02230
Event Type
Malfunction
Date Received
June 5, 2025
Date of Event
May 20, 2025
Report Date
June 30, 2025
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
UDI-DI
00763000272739
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: ANALYSIS WAS PERFORMED FOR PRODUCT: 9735821R, LOT NUMBER: P905077. IT WAS REPORTED THAT THE RETURNED POSITIONING SENSOR UNIT (PSU) HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .367 MM WITH A PASSING THRESHOLD OF .250 MM. ALREADY REPORTED CODES B01, C08 AND D02 ARE APPLICABLE TO THIS ANALYSIS. SEE ALSO SECTION E FOR ADDITIONAL INFORMATION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821R, UBD: UNKNOWN, UDI#: UNKNOWN. WORK ORDER COMPLETED: H3, H6) A MANUFACTURER REPRESENTATIVE WENT TO THE SITE TO TEST THE NAVIGATION SYSTEM. IT WAS REPORTED THAT THE FAULTED CAMERA WAS REPLACED TO RESOLVE THE ISSUE. CODES B01, C08 AND D02 ARE APPLICABLE. A05: APPLICABLE TO LOCALIZER FAULTED. A1102: APPLICABLE TO THE LOCALIZER FAULTED MESSAGE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHILE SETTING UP FOR A PROCEDURE, THE SYSTEM DISPLAYED A LOCALIZER FAULTED MESSAGE. THE PROBABLE CAUSE WAS SUSPECTED TO BE THE CMOS (COMPLEMENTARY METAL OXIDE SEMICONDUCTOR) BATTERY IN THE CAMERA. THERE WAS NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
551524 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC 9735665 00763000272739

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."