TRUSTEEL
Report
- Report Number
- 3003442380-2025-10296
- Event Type
- Malfunction
- Date Received
- June 5, 2025
- Date of Event
- May 18, 2025
- Report Date
- September 18, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018426
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION: THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF 3709030 DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF 3709030. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(6). THE BATCH: 6010737, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 17-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010737". THE COUNT OF COMPLAINT IS 0 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT: 6010737 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 99 AND MANUFACTURED IN THE MACHINE MULTIVAC 14 ON 19/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. GLUE-CONNECTOR LOT: THE LOT: 4M01487 WAS MANUFACTURED ACCORDING TO THE (WI) VERSION 65 AND MANUFACTURED IN THE MACHINE SC05 AND SC06 ON 09/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. THE LOT: 4M01490 WAS MANUFACTURED ACCORDING TO THE (WI) VERSION 65 AND MANUFACTURED IN THE MACHINE SC08 ON 10/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: ONE EXTENDED WAS RAISED DURING THE PROCESS UNRELATED TO THE MALFUNCTION REPORTED, THEREFORE, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. IT WAS REPORTED THAT PATIENT FACED INFUSION SET DETACHMENT EVENT ON (B)(6) 2025. THE SITE OF DETACHMENT WAS FROM HUB. THE INFUSION SET WAS IN USE FOR 1 DAY. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 761934 | TRUSTEEL | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002833 | 6010737 | 05705244018426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |