FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 2214540 · Received August 3, 2011

Report

Report Number
2027969-2011-01722
Event Type
Malfunction
Date Received
August 3, 2011
Date of Event
July 5, 2011
Report Date
August 3, 2011
Manufacturer
ALERE SAN DIEGO, INC
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: INRATIO: 4.3, REFERENCE: 2.84, MEAN: 3.57, CONFIDENCE LIMITS: 2.0-5.0. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT INRATIO AND REFERENCE TEST RESULT COMPARISON DID MEET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. NO FURTHER INVESTIGATION REQUIRED. PER COMPLAINT ISSUE, PATIENT HAS HYPOTHYROIDISM AND IS TAKING NORCO. PER PRODUCT USER GUIDE, "CERTAIN PRESCRIPTION DRUGS AND OVER-THE-COUNTER MEDICATIONS (E.G., ANTIBIOTICS, PAIN RELIEVERS) CAN AFFECT THE ACTION OF ORAL ANTICOAGULANTS. STARTING, STOPPING OR CHANGING THE DOSE CAN AFFECT THE INR VALUE." PATIENT CURRENT MEDICATION AND HEALTH STATUS MAY LEAD TO UNEXPECTED INR RESULT OR TESTING ERROR." (B)(4). ACTION THRESHOLD OF (B)(4) WAS REACHED. STRIP LOT #243104 IN COMPLAINT HAS STRIP CODE Z45BU WITH BRICK LOT #246544, WHICH WAS SPLIT INTO TWO DIFFERENT LOTS (243104 AND 251117). (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLD OF (B)(4), NO FURTHER ACTION IS REQUIRED. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 4.3, LAB: 2.84. PATIENT'S THERAPEUTIC RANGE: 2.0-3.0 INR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC 100071 243104

Patients

Seq Age Sex Outcome Treatment
1 NI