FDA Adverse Event Other Summary report: N

GELWEAVE STRAIGHT

MDR report key: 2214248 · Received July 27, 2011

Report

Report Number
9612515-2011-00005
Event Type
Other
Date Received
July 27, 2011
Date of Event
June 22, 2011
Report Date
July 26, 2011
Manufacturer
VASCUTEK LTD.
Product Code
DSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: MANUFACTURING AND QC RECORDS REVIEWED. GRAFT WAS EXPLANTED AND DISPOSED OF BY THE HOSPITAL. RESULTS: THERE WERE A TOTAL OF 73 UNITS IN THE BATCH. ALL GRAFTS WERE GELATIN SEALED. AFTER SEALING, 100% OF ALL GRAFTS ARE WHOLE GRAFT POROSITY TESTED. THIS INVOLVES PRESSURIZING THE DEVICE WITH AN ISOPROPANOL ALCOHOL.GLYCEROL MIXTURE AND MEASURING THE FLOW RATE. THE MAXIMUM PERMISSIBLE FLOW RATE WAS DETERMINED BY COMPARISON WITH CITRATED COWS BLOOD. THE QUALITY CONTROL AND MANUFACTURING RECORDS WERE REVIEWED AND THERE WAS NOTHING TO SUGGEST ANY PROBLEMS. THE LEAK TEST RESULTS WERE WITHIN OUR SPECIFIED QUALITY CONTROL LIMIT FOR THIS PRODUCT. ALL GRAFTS WERE DISTRIBUTED BETWEEN (B)(4) 2011. THE USAGE RATE OF THIS PARTICULAR MODEL OF GRAFT IS SUCH THAT IT MAY BE ASSUMED THE MAJORITY HAVE NOW BEEN IMPLANTED. NO OTHER REPORTS HAVE BEEN RECEIVED FOR THE OTHER UNITS. VASCUTEK WILL RESPOND TO THE SURGEON AND PROVIDE DETAILS OF OUR INVESTIGATION IN OUR FINAL REPORT.

Description of Event or Problem · 1

THE EVENT OCCURRED AT (B)(6) IN (B)(6) IN USA. NO DEATH OR SERIOUS INJURY TOOK PLACE. THE EVENT IS BEING REPORTED AS INTERVENTION WAS REQUIRED. A GELWEAVE GRAFT WAS USED AND DURING THE OPERATION AND THE DOCTOR OBSERVED A SERIOUS AMOUNT OF BLOOD LEAKAGE FROM THE GRAFT. ANOTHER GRAFT WAS USED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THE PATIENT RECOVERED WELL FROM THE OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GELWEAVE STRAIGHT VASCULAR PROSTHESIS DSY VASCUTEK LTD. GELWEAVE 120238/1 3341

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention