Malfunction
Summary report: N
TORX 8 DRIVER, NIPPLE TIP
MDR report key: 22142150
·
Received June 4, 2025
Report
- Report Number
- 2031009-2025-00012
- Event Type
- Malfunction
- Date Received
- June 4, 2025
- Date of Event
- May 7, 2025
- Report Date
- June 4, 2025
- Manufacturer
- TRIMED, INC.
- Product Code
- HXX
- UDI-DI
- 00842188126048
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- 003
Narratives
Description of Event or Problem · 0
A SALES REP REPORTED ON (B)(6) 2025 THAT WHEN A SURGEON WAS TIGHTENING THE 2ND MOST RADIAL 2.4 TPEG INTO THE VOLAR FIXED ANGLE PLATE THE TIP OF THE DVTX-8 NP BROKE FLUSH WITH THE SCREW HEAD. IT WAS REPORTED THAT THE BROKEN PIECE OF THE DRIVER TIP WAS UNABLE TO BE REMOVED FROM THE SCREW. NO CURRENT PATIENT IMPACT REPORTED OR TIME DELAY REPORTED BY THE SURGEON. POTENTIAL FOR PATIENT IMPACT IF PROSTHESIS REMOVAL IS REQUIRED IN THE FUTURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 627137 | TORX 8 DRIVER, NIPPLE TIP | TORX 8 DRIVER, NIPPLE TIP | HXX | TRIMED, INC. | DVTX-8 NP | 00842188126048 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |