FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 20GAX1.16IN PRN SLM NPVC

MDR report key: 22141796 · Received June 4, 2025

Report

Report Number
3006948883-2025-00179
Event Type
Malfunction
Date Received
June 4, 2025
Date of Event
March 10, 2025
Report Date
June 9, 2025
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830572
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW(LOT#4171666): 1)THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 4 IN JULY 2024, AND PACKAGED AT CFS PACKAGE LINE IN JULY 2024. WORK ORDER QUANTITY WAS (B)(4) EA. 2)REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. AMONG THEM, THE PRN TORQUES MEET THE OUTGOING INSPECTION REQUIREMENT. 3)REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLE AND PHOTO HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. CHECK THE RETAINED SAMPLES OF THIS BATCH, NO FOREIGN MATTERS ARE FOUND. 4. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITIES ARE FOUND IN THE PROCESS AND RETAINED SAMPLES, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS. AS THE STATUS AND COMPOSITION OF THE BLACK SPOT ON THE INDWELLING NEEDLE CANNOT BE CONFIRMED, SO THE ROOT CAUSE OF THIS DEFECT CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO PAY ATTENTION TO AND MONITOR SUCH DEFECTS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II 20GAX1.16IN PRN SLM NPVC HAD FOREIGN MATTER. THE INDWELLING NEEDLE HAS A BLACK SPOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
529995 BD INTIMA-II 20GAX1.16IN PRN SLM NPVC INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 4171666 00382903830572

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown