FDA Adverse Event Other Summary report: N

TISSUE RETRIEVAL BAG

MDR report key: 2214102 · Received July 18, 2011

Report

Report Number
1416891-2011-00005
Event Type
Other
Date Received
July 18, 2011
Date of Event
May 9, 2011
Report Date
July 15, 2011
Manufacturer
ANCHOR PRODUCTS CO.
Product Code
GCJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

NO SAMPLE WAS RETURNED. ATTEMPTS TO TEAR THE BAG ALONG THE BASE COULD NOT BE DUPLICATED. A PULL TEST STUDY WAS CONDUCTED TO THE POINT OF BREAKAGE ON A SAMPLE OF BAGS AND RESULTS AVERAGED 29.5 LBF. THE METHOD USED TO ADVANCE THE BAG THROUGH THE TROCAR IS NOT INDICATED ON THE REPORT, RETRIEVAL BAGS CAN BE DAMAGED IF PASSED THROUGH STAINLESS STEEL REUSABLE TROCARS WHICH HAVE NOT BEEN ADEQUATELY MAINTAINED. SHARP EDGES ON THE ENTRANCE POSITION CAN DAMAGE THE BAG MATERIAL. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

NOTE THAT THE DATE OF EVENT WAS RECORDED AS (B)(6) 2011, BUT WAS NOT REPORTED TO ANCHOR UNTIL (B)(6) 2011. BOTTOM SECTION OF TISSUE RETRIEVAL BAG (ALONG THE BASE) WAS REPORTED TO HAVE SHEARED OFF DURING INSERTION OF THE BAG INTO THE TROCAR. THE BASE OF THE BAG FELL INTO THE ABDOMEN BUT WAS RETRIEVED BY THE SURGEON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TISSUE RETRIEVAL BAG TRS080 GCJ ANCHOR PRODUCTS CO. TRS080 R62N

Patients

Seq Age Sex Outcome Treatment
1