ETRIO
Report
- Report Number
- 3005075853-2011-03347
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Date of Event
- July 26, 2011
- Report Date
- July 29, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K072493
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. THE COMPLAINT WAS FOR AN (B)(4) BUT RECEIVED WAS AN (B)(4). THE DEVICE WAS TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED AND THE CUTTING OF THE TEST MEDIA PERFORMED AS EXPECTED. NO ISSUES WERE FOUND WITH THE HANDLE. THE INSTRUMENT WOULD OPEN AND CLOSE AS EXPECTED. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. IT IS POSSIBLE THAT THE DEVICE RETURNED WAS THE DEVICE USED TO COMPLETE THE PROCEDURE.
IT WAS REPORTED THAT DURING A COLON RESECTION THE DEVICE WOULD NOT OPEN WHEN THE HANDLE WAS ACTUATED. THE DEVICE WAS NOT ON A VESSEL OR ARTERY WHEN THE JAWS WOULD NOT OPEN. THERE WAS NO TISSUE TRAUMA REPORTED. THEY COMPLETED THE PROCEDURE WITH A NEW LIKE DEVICE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETRIO | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | UNK | H4437G |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |