FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2213904 · Received August 17, 2011

Report

Report Number
6000001-2011-19327
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
August 3, 2011
Report Date
August 3, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION WHERE THE PUMP "DID NOT PASS CALIBRATION PROCEDURE FOR MINIMUM DOWNSTREAM OCCLUSION PRESSURE" WAS NOT CONFIRMED OR REPRODUCED DURING PRODUCT EVALUATION. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THIS CONDITION AND NO REPAIR WAS NECESSARY. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION OF "FAILED TEST (DOWNSTREAM OCCLUSION SENSOR CALIBRATION)" ISSUES. DURING PREVIOUS SERVICE ON (B)(6) 2006 FOR (B)(4) THE PHM WAS REPLACED. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP THAT DID NOT PASS CALIBRATION PROCEDURE FOR MINIMUM DOWNSTREAM OCCLUSION PRESSURE. THIS MALFUNCTION COULD CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY WERE IT TO RECUR. THIS CONDITION WAS FOUND IN THE BIOMED DEPARTMENT UPON POWER UP. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1