FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TEST STRIPS
MDR report key: 2213841
·
Received August 4, 2011
Report
- Report Number
- 1826988-2011-00448
- Event Type
- Malfunction
- Date Received
- August 4, 2011
- Date of Event
- July 1, 2011
- Report Date
- July 12, 2011
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- PMA / PMN Number
- K091820
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER TESTED HER BLOOD GLUCOSE USING HER 2 CONTOUR USB METERS AND RECEIVED READINGS OF "LOW" AND 80 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. REPLACEMENT STRIPS AND A NEW METER WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 7080G | 0KC3C03 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |