VENTED AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2011-00501
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- July 10, 2011
- Report Date
- July 22, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE RETURNED MR290V CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE AND WAS CONNECTED TO A WATERBAG FOR FUNCTIONAL TESTING. RESULTS: THE CHAMBER FILLED NORMALLY WHEN CONNECTED TO A WATERBAG; HOWEVER, ONCE THE CHAMBER HAD FILLED, SMALL DROPS OF WATER WERE OBSERVED TO LEAK FROM THE CONNECTION BETWEEN THE FEEDSET TUBE AND WATERBAG SPIKE. INSPECTION SHOWED THAT INSUFFICIENT GLUE HAD BEEN APPLIED BETWEEN THE CONNECTION OF THE FEEDSET TUBE AND SPIKE, CAUSING THE LEAK. A LOT CHECK REVEALED SIX OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100222. CONCLUSION: ALL CHAMBERS ARE PRESSURE TESTED BEFORE LEAVING THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." THE WATERFEED TUBING IS ATTACHED TO THE SPIKE BY AN AUTOMATED GLUING PROCESS. AS PART OF OUR ONGOING IMPROVEMENT PROCESS, ADDITIONAL GLUE IS NOW APPLIED TO THE WATER FEEDSET SPIKE DURING THE AUTOMATED ASSEMBLY. (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT WATER LEAKED FROM THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND SPIKE OF TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS NOTICED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED HUMIDIFICATION CHAMBER | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | MR290V | 100222 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RT204 ADULT DUAL-HEATED BREATHING CIRCUIT |