FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 2213755 · Received August 16, 2011

Report

Report Number
9611451-2011-00501
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 10, 2011
Report Date
July 22, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED MR290V CHAMBER WAS VISUALLY INSPECTED FOR DAMAGE AND WAS CONNECTED TO A WATERBAG FOR FUNCTIONAL TESTING. RESULTS: THE CHAMBER FILLED NORMALLY WHEN CONNECTED TO A WATERBAG; HOWEVER, ONCE THE CHAMBER HAD FILLED, SMALL DROPS OF WATER WERE OBSERVED TO LEAK FROM THE CONNECTION BETWEEN THE FEEDSET TUBE AND WATERBAG SPIKE. INSPECTION SHOWED THAT INSUFFICIENT GLUE HAD BEEN APPLIED BETWEEN THE CONNECTION OF THE FEEDSET TUBE AND SPIKE, CAUSING THE LEAK. A LOT CHECK REVEALED SIX OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100222. CONCLUSION: ALL CHAMBERS ARE PRESSURE TESTED BEFORE LEAVING THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." THE WATERFEED TUBING IS ATTACHED TO THE SPIKE BY AN AUTOMATED GLUING PROCESS. AS PART OF OUR ONGOING IMPROVEMENT PROCESS, ADDITIONAL GLUE IS NOW APPLIED TO THE WATER FEEDSET SPIKE DURING THE AUTOMATED ASSEMBLY. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT WATER LEAKED FROM THE CONNECTION BETWEEN THE WATER FEEDSET TUBE AND SPIKE OF TWO MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS NOTICED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTED AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 100222

Patients

Seq Age Sex Outcome Treatment
1 RT204 ADULT DUAL-HEATED BREATHING CIRCUIT