BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2025-00975
- Event Type
- Malfunction
- Date Received
- June 3, 2025
- Date of Event
- May 10, 2025
- Report Date
- June 12, 2025
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- UDI-DI
- 30382903690320
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY BD HAD NOT RECEIVED ANY SAMPLES FOR INVESTIGATION. A TOTAL OF 200 RETAINED SAMPLES, 100 FROM EACH LOT NUMBER, WERE VISUALLY INSPECTED, AND NO ADDITIVE ABNORMALITIES WERE FOUND. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF MAY 2025. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOTS, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE LOT 4100736, FOR THE INDICATED FAILURE MODE: HEMOLYSIS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.
REPORT 1 OF 2. IT WAS REPORTED WHILE USING BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED HEMOLYSIS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
REPORT 1 OF 2. IT WAS REPORTED WHILE USING BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES EXHIBITED HEMOLYSIS. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 362178 | BD VACUTAINER® CAT (CLOT ACTIVATOR TUBE) PLUS BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | 4100736 | 30382903690320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |