FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2213682 · Received August 16, 2011

Report

Report Number
6000001-2011-19259
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 24, 2011
Report Date
August 2, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH THE DISPLAY "GOES LIGHT AND DARK" WAS CONFIRMED AND REPRODUCED BY SERVICE. THIS CONDITION WAS CAUSED BY A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). THE UIM PCB WAS REPLACED TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WHERE THE DISPLAY GOES LIGHT AND DARK; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND UPON POWER UP. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1