COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-19259
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- July 24, 2011
- Report Date
- August 2, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH THE DISPLAY "GOES LIGHT AND DARK" WAS CONFIRMED AND REPRODUCED BY SERVICE. THIS CONDITION WAS CAUSED BY A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). THE UIM PCB WAS REPLACED TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WHERE THE DISPLAY GOES LIGHT AND DARK; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND UPON POWER UP. THERE WAS NO REPORTED PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SOFTWARE VERSION IS CURRENTLY UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |