INFANT BIAS FLOW DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00497
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Report Date
- July 21, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K034026
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE RETURNED COMPLAINT BREATHING CIRCUIT WAS VISUALLY INSPECTED AND PRESSURE TESTED. THE VISUAL INSPECTION REVEALED NO PHYSICAL DAMAGE TO THE COMPLAINT DEVICE. THE PRESSURE TEST REVEALED THE PRESSURE DROP TO BE WITHIN SPECIFICATION FOR THIS PRODUCT. NO FAULT WAS FOUND WITH THE RETURNED COMPLAINT BREATHING CIRCUIT. ALL CIRCUITS ARE PRESSURE AND FLOW TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCTION LINE. THIS IS (B)(4) ANY THAT FAIL THESE TESTS ARE DISCARDED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT235 STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS."
A HOSPITAL IN (B)(6) REPORTED THAT THE SWIVEL CONNECTOR OF AN RT235 INFANT BIAS FLOW DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS "LEAKY". THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT BIAS FLOW DUAL-HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT235 | 110406 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |