FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2213634 · Received August 16, 2011

Report

Report Number
2531779-2011-05958
Event Type
Malfunction
Date Received
August 16, 2011
Report Date
July 20, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION CONFIRMED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING DESIRED PUMP FUNCTIONS WHEN PRESSED. THE BUTTONS DO NOT SPRING BACK OR CLICK NORMALLY. CONTAMINATION WAS FOUND UNDER THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT STATED THAT THE UP, DOWN, AND OK BUTTONS REQUIRE MULTIPLE PRESSES TO ACTIVATE A PUMP RESPONSE. THE HIGHLIGHT ON THE PUMP MENU SCREENS THEN SCROLL WHEN THE BUTTON ENGAGES AND THE BUTTONS CLICK/SPRING BACK NORMALLY. THE UP ARROW BUTTONS REPORTEDLY FEELS "MUSHY" AND LOOSE. THE PATIENT DOES NOT CLEAN THE PUMP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 55 YR