FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2213615 · Received August 16, 2011

Report

Report Number
6000001-2011-19210
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
June 30, 2011
Report Date
July 28, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH A MALFUNCTION OF "DOOR OPEN CLOSE CLAMP" WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY A BROKEN DOOR. THE PUMP HEAD DOOR AND REQUIRED PARTS WERE REPLACED TO CORRECT THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT OF A FLO-GARD INFUSION PUMP WITH A DOOR OPEN ALARM, WHICH COULD HAVE CAUSED AN INTERRUPTION OF DELIVERY. THE REPORTED CONDITION OCCURRED UPON POWER UP IN THE GENERAL PATIENT WARD. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1