FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2213604 · Received August 16, 2011

Report

Report Number
6000001-2011-19200
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
June 30, 2011
Report Date
July 26, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE PUMP WITH A MALFUNCTION OF A DAMAGED BATTERY WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY, WHICH COULD HAVE CAUSED AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90 CATEGORIZED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1