FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM

MDR report key: 2213568 · Received August 16, 2011

Report

Report Number
2050012-2011-04535
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 17, 2011
Report Date
July 17, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CHECKED THE PROBE TUBING, THE SYRINGE, AND THE TUBING TO THE VALVE. THE CUSTOMER STOPPED, HOMED THE SYSTEM, AND RAN QC. THE DRIPPING FROM THE PROBE CONTINUED. ON (B)(4) 2011, BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT. THE FSE FOUND THE REAGENT PROBE WASH COLLAR LEAKING, AND REAGENT A VACUUM VALVE WAS WORN OUT. THE FSE REPLACED THE VACUUM VALVE. THE FSE VERIFIED THE PROBES WERE CLEAR, AND REALIGNED WASH STATION. THE FSE PERFORMED CUVETTE WASH WITH NO ERRORS OR LEAKING OBSERVED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE CARTRIDGE CHEMISTRY SAMPLE PROBE ON UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEM WAS DRIPPING FLUID. NO PATIENT SAMPLE WAS RUN ON THE INSTRUMENT ON THE DAY OF THE EVENT. THE CUSTOMER WAS WEARING PPE. MSDS WAS NOT REVIEWED, BUT THE FACILITY HAS AN EXPOSURE CONTROL PLAN. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION AND THERE WAS NO EFFECT TO PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER, INC. DXC 600 PRO NA

Patients

Seq Age Sex Outcome Treatment
1