FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2213521 · Received August 16, 2011

Report

Report Number
2050012-2011-04275
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 17, 2011
Report Date
July 19, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS SERUM. NO ADDITIONAL INFORMATION AS PROVIDED. PER THE CUSTOMER, QC RUN PRIOR TO THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES. QA REVIEW OF THE QC SHOWS THAT THE LAB'S RANGES ARE WIDE, AND QC DID EXHIBIT IMPRECISION IN THE DAYS PRIOR TO THE EVENT. A BEC FIELD SERVICE ENGINEER (FSE) FOUND A BUBBLE ON THE CO2 MEASURING ELECTRODE. THE FSE REPLACED THE ELECTRODE AND MEMBRANE AND VERIFIED PERFORMANCE.

Additional Manufacturer Narrative · 1

CHANGE FROM: THE SAMPLE WAS SERUM. NO ADDITIONAL INFORMATION AS PROVIDED. TO: THE SAMPLE WAS SERUM, COLLECTED IN SST TUBES AND LITHIUM HEPARIN TUBES AND SPUN DOWN FOR 10 MINUTES.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) IN REGARDS TO INSTRUMENT GENERATED INTERMITTENT ERRONEOUS CO2 RESULTS. ERRONEOUS RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THE SAMPLE WAS REPEATED ON THE ALTERNATE DXC IN THE LAB AND THE OBTAINED RESULT WAS REPORTED. PATIENT RESULTS ARE PROVIDED. PATIENT TREATMENT WAS NOT IMPACTED BY THIS EVENT

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 600I N/A

Patients

Seq Age Sex Outcome Treatment
1 87 YR