FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2213426 · Received August 16, 2011

Report

Report Number
2531779-2011-05949
Event Type
Malfunction
Date Received
August 16, 2011
Report Date
July 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVALUATION CONFIRMED THAT THE KEYPAD BUTTONS DO NOT ACTIVATE DESIRED PUMP FUNCTIONS WHEN PRESSED. CONTAMINATION AND CORROSION WAS FOUND UNDER THE BOLUS BUTTON CONNECTOR LEADS. AFTER THE BOLUS BUTTON WAS REMOVED THE OTHER KEYPAD BUTTONS FUNCTIONED PROPERLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT NOTICED THAT KEYPAD BUTTONS WERE NOT ACTIVATING DESIRED PUMP FUNCTIONS WHEN PRESSED. THE PATIENT DENIES CLEANING THE PUMP. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 27 YR