Report Number
9610773-2025-03721
Event Type
Malfunction
Date Received
June 3, 2025
Date of Event
May 7, 2025
Report Date
August 28, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761074971
PMA / PMN Number
K190744
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: D8, G1, H2, H3, H6, H11. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND THE REPORTED FAILURE LIGHT BROKEN WAS CONFIRMED AND THE REPORTED FAILURE SLIGHT SCRATCH ON TIP, DEW ON LENS WAS NOT CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: THE LG-BUNDLE OF THE VIDEO CABLE SECTION IS BROKEN, THE LIGHT TRANSMISSION IS LOST OR TOO LOW. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE CAUSE FOR THE REPORTED EVENT SLIGHT SCRATCH ON TIP, DEW ON LENS COULD NOT BE DETERMINED AND FOR THE ADDITIONAL REPORTABLE FINDINGS, THE MOST PROBABLE CAUSE TRACED TO COMPONENT FAILURE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE RIGID VIDEO SCOPE HAD LIGHT BROKEN, DEW ON LENS AND SLIGHT SCRATCH ON TIP. THERE WERE NO REPORTS OF PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
888041 VIDEO TELESCOPE "ENDOEYE HD II", 10 MM, 30°, AUTOCLAVABLE LAPAROSCOPE, GYNECOLOGIC HET OLYMPUS WINTER & IBE GMBH WA50042A 04042761074971

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown