FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2213328 · Received August 16, 2011

Report

Report Number
6000001-2011-19113
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
May 29, 2011
Report Date
July 26, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE IS UNKNOWN. NO REPAIRS WERE MADE TO CORRECT THE REPORTED CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A FAILURE CODE 812:05, WHICH INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1