FDA Adverse Event
Malfunction
Summary report: N
FLEXCATH STEERABLE SHEATH
MDR report key: 2213299
·
Received August 16, 2011
Report
- Report Number
- 3002648230-2011-00130
- Event Type
- Malfunction
- Date Received
- August 16, 2011
- Date of Event
- July 19, 2011
- Report Date
- July 19, 2011
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
Description of Event or Problem · 1
THERE WAS FLUID LEAKING FROM THE VALVE OF THE FLEXCATH STEERABLE SHEATH. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 | 61555 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |