FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2213251 · Received August 16, 2011

Report

Report Number
2531779-2011-05942
Event Type
Malfunction
Date Received
August 16, 2011
Date of Event
July 19, 2011
Report Date
July 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4): TESTING CONFIRMED THE DOWN BUTTON IS UNRESPONSIVE. THE KEYPAD WAS TORN AND WAS REMOVED FOR INVESTIGATION. CONTAMINATION WAS FOUND UNDER ALL BUTTON CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE DOWN ARROW BUTTON STOPPED RESPONDING. THE PATIENT CONFIRMED THERE WERE NO HOLES OR TEARS IN THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP ON THE OUTSIDE OF CLOTHING AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 40 YR