FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUPS 58

MDR report key: 2213235 · Received August 12, 2011

Report

Report Number
1818910-2011-14939
Event Type
Injury
Date Received
August 12, 2011
Report Date
February 12, 2014
Manufacturer
DEPUY INTERNATIONAL, LTD.
Product Code
KWA
PMA / PMN Number
PK040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 1

THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4).DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

LITIGATION PAPERS ALLEGE: PATIENT WAS IMPLANTED WITH A DEPUY ASR HIP ON HIS LEFT SIDE ON OR ABOUT (B)(6) 2007. PATIENT WAS IMPLANTED WITH A DEPUY ASR HIP ON HIS RIGHT SIDE ON OR ABOUT (B)(6) 2008. SINCE HIS SURGERIES, PATIENT HAS EXPERIENCED PAIN. HE HAS NOT YET SCHEDULED AN EXPLANTATION OF EITHER ASR HIP IMPLANT.

Description of Event or Problem · 1

**UPDATE** (B)(4) 2013 - SALES REP REPORTED REVISION SURGERY. PATIENT WAS REVISED TO ADDRESS ELEVATED METAL ION LEVELS.

Description of Event or Problem · 1

UPDATE: 2/12/2014 - MEDICAL RECORDS RECEIVED FOR RIGHT HIP REVISION SURGERY. PATIENT'S RIGHT HIP WAS REVISED TO ADDRESS A CYST, FIBRINOUS DEBRIS, BONE EROSION, GRANULAR TYPE MATERIAL AND METALLOSIS. THE INFORMATION RECEIVED DOES NOT CHANGE THE MDR DECISION. THIS COMPLAINT WAS UPDATED ON: 02/25/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASR ACETABULAR CUPS 58 TOTAL HIP REPLACEMENT KWA DEPUY INTERNATIONAL, LTD. NA 2557520

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention