FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2213129 · Received August 16, 2011

Report

Report Number
2531779-2011-05932
Event Type
Injury
Date Received
August 16, 2011
Date of Event
July 17, 2011
Report Date
July 17, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. REVIEW OF THE PUMP HISTORY REVEALED 9 BOLUS DOSES GIVEN WITHIN A 9-MINUTE PERIOD OF TIME THAT MAY HAVE CONTRIBUTED TO THE PATIENT'S LOW BLOOD GLUCOSE EVENT. THERE WAS NO EVIDENCE OF ANY DEVICE MALFUNCTION. THE ANIMAS REPRESENTATIVE RECOMMENDED THAT ADVANCED BOLUS FEATURES BE REVIEWED WITH A HEALTH CARE PROVIDER.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4) 2011. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.

Description of Event or Problem · 1

THE PATIENT'S WIFE STATED THAT THE PATIENT'S BLOOD GLUCOSE FELL TO 20 MG/DL AND THAT THE PATIENT LOST CONSCIOUSNESS. REVIEW OF THE PUMP HISTORY REVEALED THAT THE TOTAL DAILY DOSE FOR THE DAY HE LOST CONSCIOUSNESS WAS 60 UNITS. THE PATIENT AVERAGES BETWEEN 30 AND 35 UNITS OF INSULIN DELIVERED PER DAY. REVIEW OF THE BOLUS HISTORY REVEALED 9 BOLUS DOSES OF 4 UNITS EACH PROGRAMMED WITH THE NORMAL BOLUS FEATURE BETWEEN 12:33 AND 12:42 PM. THE PARAMEDICS WERE CALLED AND GLUCAGON WAS ADMINISTERED TO CORRECT THE LOW BLOOD GLUCOSE LEVEL. THE PATIENT'S BLOOD GLUCOSE LEVEL INCREASED TO 167 MG/DL. THE PATIENT REPORTEDLY HAD SOMETHING TO EAT AFTERWARD ALTHOUGH THE WIFE DID NOT SPECIFY WHAT HE HAD TO EAT. WHEN THE PARAMEDICS LEFT HIS BLOOD GLUCOSE LEVEL WAS REPORTEDLY 122 MG/DL. HIS BLOOD GLUCOSE LEVEL BEFORE DINNER WAS 70 MG/DL WITHOUT ANY SYMPTOMS.