FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2213115 · Received August 16, 2011

Report

Report Number
2531779-2011-05936
Event Type
Malfunction
Date Received
August 16, 2011
Report Date
July 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

SHOULD HAVE READ: "(B)(4) DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS:". THE PRODUCT WAS INVESTIGATED ON (B)(4) 2012.

Additional Manufacturer Narrative · 1

(B)(4): ALL OF THE KEYPAD BUTTONS WERE FOUND TO BE RESPONDING APPROPRIATELY TO BUTTON PRESSES. THE KEYPAD WAS REMOVED AND NO BUTTON CONTACT ISSUES WERE FOUND.

Description of Event or Problem · 1

THE REPORTER STATED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING DESIRED PUMP FUNCTIONS WHEN PRESSED. THE ISSUE APPEARED TO BE OCCURRING AFTER MULTIPLE BUTTONS. THE PATIENT REPORTEDLY CLEANS THE PUMP WITH AN ANTIBACTERIAL WIPE. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1 7 YR